OD600
(Cell Density, Bacterial Growth, Yeast Growth)
The OD600 Basics
Optical Density (OD) measurements of microbial and cell growth are one of the most common methods used in a microbiology lab. Some of the main applications are the determination of the optimal time at which to harvest, the determination of the optimal time to induce a culture when running a protein expression protocol or the monitoring of cloning procedures.
The growth of cells, bacteria or yeast (cell density, bacterial growth, yeast growth) in liquid culture media is commonly controlled by measuring the optical density at 600 nm (OD600). OD600 measurements are typically used to determine the stage of growth of a bacterial culture, these measurements help ensure that cells are harvested at an optimum point that corresponds to an appropriate density of live cells.
Since optical density in the case of OD600 measurements results from light scattering rather than light absorption, size and shape as well as dead cells and debris of a cell may add to light dissipating. Distinctive cell types that are at densities of the same level (eg. cells/mL) may therefore show varying values OD600 when estimated on a similar instrument. Growth of bacterial cells typically progresses through a series of consecutive phases including lag, log, stationary, and decline.
In general, cells should be harvested towards the end of the log phase, using the optical density of the samples to determine when this point has been reached. Cells are routinely grown until the absorbance at 600 nm (known as OD600) reaches approximately 0.4 prior to induction or harvesting.
Bacterial growth curve
The incoming light is emitted by the cells/particles in the sample randomly. This results in an outgoing stray light in undefined direction.
Light scattering of turbid sample
The incoming light is reduced by sample absorption (outgoing light) depending on sample concentration and wavelength. The Beer-Lambert low describes the correlation of incoming/outgoing light and sample concentration.
Light absorption of clear sample
The Beer-Lambert Law and OD600
The Lambert-Beer-Law, also known as Beer’s Law, empirically relates the absorption of light to the properties of the sample. This law states that there is a logarithmic relationship between the transmission of light through a specific sample (T = I/Io with I = outgoing light and Io = incoming light), the molar extinction coefficient for a specific compound (ε), the concentration of the absorbing species in the material (c) and the distance the light travels (d).
OD estimations have gotten synonyms with estimations of bacterial concentration (C) or number (N), as per the Beer-Lambert law. However measurements of OD are actually measurements of turbidity, we can therefore apply the Beer-Lambert law, with certain contemplations, just for microbial colonies of low densities.
Difference in OD600 Readings
For turbid samples such as cell cultures, the major contributor for the absorbance measured is light scattering and not the result of molecular absorption following the Beer-Lambert Law. The measurements are therefore depending on the optical setup of the spectrophotometer (distance between the cell holder and instrument exit slit, monochromator optics, slit geometry, etc.), different instrument types will most likely tend to give different OD 600 readings for the same turbid sample.
Therefore, if results from different spectrophotometers are to be compared, they must be normalized first by either a simple correlation of the OD readings using the same sample on the two different instruments to compare
or with a more precise approach by the creation of appropriate calibration curves.
A calibration curve can be constructed by comparing measured OD 600 to expected OD 600 over a range of different concentrations. Expected OD 600 is determined by counting cell number using an alternative technique (for example microscope slide method) and converting to OD 600 using the rule of thumb that 1 OD 600 = 5 x 108 cells/ml for E. coli. A high-quality instrument like the NanoPhotometer® or the OD600 DiluPhotometer™ comes with a “correction factor” of 1 as default, which, once adjusted will help to provide comparable results between different equipment.
The use of disposable cuvettes is recommended rather than micro/nano volume technologies (measurements in a drop) for optical density measurements of cell culture solutions. The amount of cells is reflected in the reading and the likelihood of fluctuating amount of cells in a drop from sample to sample can be considered as extremely significant.
It is therefore recommended to use cuvettes since the amount of error in a bigger volume is not as significant. The cuvette measurements provide a bigger average and therefore more reproducible readings. Also, to prevent the suspension settling too quickly and giving an OD reading that changes with time, glycerol should be added to the sample.
Higher absorbance reading
Lower absorbance reading
Linear Range for OD600 Measurements
The linear range of the instrument used for the measurements is critical to cover the entire growth cycle of the culture, an OD limit of at least 2.5 OD is advised to provide sufficient working range. Please bear in mind that the sample may not be linear over the entire range if a high amount of dead cells are present, obscuring the result. If samples should still be out of range, dilutions may become necessary; the parameters should be carefully selected to cover the entire desired area without compromising the linearity of the setup.
To reduce the required volume for the OD600 measurement and to avoid time-consuming and error-prone manual dilutions, special cuvettes are available for the researcher. DiluCell™ allows for lower sample volume requirements and provides automatic virtual sample dilution. Due to the unique design of DiluCell™ cuvettes, the light-path is reduced from the standard 10 mm to 0.5 mm (DC 20) or 1.0 mm (DC 10). According to the Lambert Beer Law, this shortened pathlength results in an automatic sample dilution by factor 20 (DC 20) or 10 (DC 10). Automatic dilution with DiluCell™ saves time and excludes dilution errors and cross contaminations.
OD600 Tool
Click on the link below to access a spreadsheet OD600 Tool, version 1.2, that helps generate microbial growth curves according to OD600 values as well as to calculate other values of interest for microbiology experiments like cell forming units CFU/mL or time of inoculation. Once instrument calibration and standard curve have been established, it can easily be used to document growth experiments for any type of cell culture.
Implen GmbH
Implen, based in Munich, Germany, is a leading supplier for innovative spectroscopy instruments and consumables for the non-destructive analysis of liquid samples offering a complete range of spectrophotometers for cuvette and drop measurements.
Visit www.implen.de to find out how DiluPhotometer® and NanoPhotometer® can improve your research.
Spectrophotometers and accessories for OD600 readings
The temperature-controlled cuvette port of the NanoPhotometer® C40 provides an unmatched dynamic range of up to 2.6 Abs without the need for a cover during the reading. OD600 readings for Cell Density, Bacterial Growth and Yeast Growth are possible much faster and within a comfortable dynamic range. Features rapid scan capability for 200 – 900 nm.
DiluCell™ provides automatic 1/10 and 1/20 dilutions of higher concentrated samples and bubble-free filling for standard 1 cm cuvette compatible spectrophotometers and is the ideal tool for OD600 readings for Cell Density, Bacterial Growth and Yeast Growth. DiluCell™ allows for convenient spectrophotometric analysis from 340 – 1100 nm.
DiluPhotometer™ OD600 is a small, easy-to-use instrument dedicated to measuring samples near the 600 nm wavelength (OD600). It is suitable for measuring growth rates of several cell types (cell density, bacterial growth, yeast growth) including E. coli and yeast and also provides software to run Bradford Assays for the analysis of proteins.